Force and displacement performance should be specified for the selected machine, measuring function and usable range—not assumed to be identical across every furniture tester. Confirm the proposed error limits, resolution, operating conditions and verification evidence before ordering. A sensor label or display increment alone does not establish the performance of the complete measurement system.
Define the Measurement You Need
State what you need to measure: force applied to the specimen, actuator movement, specimen deformation, mattress height or another quantity. Actuator travel and specimen deformation are not necessarily the same measurement. Also distinguish measurement error from how closely the controller holds the requested load during a test.
Provide the required method, specimen dimensions and expected working values if known. If the force range is unknown, describe the material or furniture and the result you need; do not select a sensor only by its maximum capacity.
Compare Proposals on the Same Basis
| Review item | What to confirm |
|---|---|
| Measuring function and range | Which quantity and usable range the declared performance applies to. |
| Performance statement | Whether an error limit is stated relative to the reading, full scale or another defined basis. |
| Resolution | The display or recorded increment, separately from measurement performance. |
| Configuration and conditions | The installed sensor, fixtures, measurement location and relevant operating conditions. |
| Evidence | The proposed verification or calibration scope, reported results and documentation included in the order. |
Check the Whole Measurement Chain
A load cell specification does not cover every effect of mounting, alignment, signal conditioning or the test setup. Likewise, an encoder reading does not automatically measure specimen deformation. Ask for evidence relevant to the installed configuration and the quantities used by your laboratory.
Use the applicable third-party metrological verification requirements as the acceptance basis for the agreed measuring functions and ranges. The required scope, conditions and acceptance criteria should be identified before verification. Review the resulting report for the actual equipment configuration. Equipment purchase and a calibration document do not, by themselves, establish laboratory accreditation or product conformity.
For terminology, consult the JCGM metrology guides and vocabulary. For purchasing records, use the furniture testing equipment calibration checklist.